location: Current position: Home >> Scientific Research >> Paper Publications

Characteristics of Magnetic Torque of a Capsule Micro Robot Applied in Intestine

Hits:

Indexed by:期刊论文

Date of Publication:2009-05-01

Journal:IEEE TRANSACTIONS ON MAGNETICS

Included Journals:SCIE、EI

Volume:45

Issue:5

Page Number:2128-2135

ISSN No.:0018-9464

Key Words:Capsule micro robot; equivalent magnetic charge method; magnetic torque; spiral drive

Abstract:A wireless drive and control method using rotational magnetic field on a capsule micro robot is proposed, and both outer and inner actuators are Nd-Fe-B magnet cylinders with multiple magnetic poles in tegular shape alternating with dissimilar radial magnetization, through magnetic coupling with rotational magnetic field generated by the rotating outer actuator, the capsule robot is rotated by an embedded inner actuator, and axial thrust is generated by fluid dynamic pressure distributing on outer surface of the robot with spiral blade, thus the screw motion of the micro robot in intestine is realized. Mathematical model on magnetic torque is established based on equivalent magnetic charge method, and the characteristics of magnetic torque with respect to structural parameters of actuators and swimming effect of the robot are studied theoretically and experimentally.

Pre One:肠道内可变直径胶囊机器人的动态特性

Next One:对培养"提高生产实习质量的探索与实践"的研究