Tube-cantilever double resonance enhanced fiber-optic photoacoustic spectrometer

点击次数:

论文类型:期刊论文

发表时间:2020-03-01

发表刊物:OPTICS AND LASER TECHNOLOGY

收录刊物:EI、SCIE

卷号:123

ISSN号:0030-3992

关键字:Trace gas detection; Double resonance; Cantilever; Fiber-optic sensor; Photoacoustic spectroscopy

摘要:An ultra-high sensitive trace gas detection method based on tube-cantilever double resonance enhanced fiberoptic photoacoustic spectroscopy (PAS) is proposed. The first-order resonant frequencies of the acoustic resonant tube and the fiber-optic cantilever microphone were both equal to the frequency of the photoacoustic pressure signal. This method combines the amplitude amplification of the photoacoustic pressure wave in an acoustic resonant tube with the response enhancement of the photoacoustic signal by the cantilever, making the gas detection extremely sensitive. An experimental double resonance enhanced photoacoustic spectrometer was built for trace acetylene detection at the wavelength of 1532.83 nm. A noise equivalent detection limit (1 sigma) was achieved to be 27 ppt with a 200-s averaging time, which is the best value reported so far. In addition, the normalized noise equivalent absorption (NNEA) coefficient reached 4.2 x 10(-10) cm(-1) W Hz(-1/2).