Hits:
Date of Publication:2011-01-01
Journal:功能材料
Volume:42
Issue:10
Page Number:1765-1767
ISSN No.:1001-9731
Abstract:We successfully prepared a PVP hydrophilic coating on medical silicon
rubber surface by appropriate treatment and silane coupling agent.The
surface morphology and the thickness of coating were investigated by
SEM(scanning electron microscope), and hydrophilic properties of the
samples were characterized by water contact angle.The tendency of
hydrophilic coating in water was observed by measuring water contact
angle of different samples with soaking time in 100mL water at 37℃.The
results indicated that the hydrophilicity of the sample treated by
coupling agent and hydrophilic solution was modified significantly.The
obtained coating was very smooth(the thickness was about 110mum), and
the hydrophilicity soaking of modified samples in water 7d was much
better than that of the matrix.
Note:新增回溯数据