陈硕
开通时间:..
最后更新时间:..
点击次数:
论文类型:期刊论文
发表时间:2009-03-15
发表刊物:ENVIRONMENTAL SCIENCE & TECHNOLOGY
收录刊物:SCIE、EI、PubMed、PKU、ISTIC、Scopus
卷号:43
期号:6
页面范围:1935-1939
ISSN号:0013-936X
摘要:A simple, environmentally friendly and continuous flow method was developed for the determination of COD based on a flow injection analysis (FIA) system, in which a BDD electrode was employed as the detecting element. The structure and the electrochemical behavior of BDD were investigated by a scanning electron microscope, Raman spectroscopy, and cyclic voltammetry, respectively. The results demonstrated that the high-quality BDD film prepared here was suitable to be used as an electrode, with which the COD measurement could be conducted. The effect of several important experimental parameters, such as applied potentials, pH, flow rates, and supporting electrolyte concentrations, on the analytical performance was investigated. Under optimized testing conditions, the proposed method was successfully applied in the COD analysis of synthetic samples. The linear range and the detection limit were 2-175 and 1 mg L(-1), respectively. In addition, the COD values determined by the proposed method compared well with those analyzed by the conventional method as demonstrated by small relative errors.