徐征
开通时间:..
最后更新时间:..
点击次数:
论文类型:期刊论文
发表时间:2016-02-01
发表刊物:MICROSYSTEM TECHNOLOGIES-MICRO-AND NANOSYSTEMS-INFORMATION STORAGE AND PROCESSING SYSTEMS
收录刊物:EI、SCIE、Scopus
卷号:22
期号:2
页面范围:283-286
ISSN号:0946-7076
摘要:In this paper, a method of electrokinetic concentrating and a glass nanofluidic device for sensitivity-enhancing of magnetic beads-based antigen-antibody immunoassay (MBAAI) are presented. In the nanofluidic device, a polyacrylamide gel plug is deployed as the nanopores via photopolymerization reaction. To verify the capability of electrokinetic concentrating, the fluorescein isothiocyanate-bovine serum albumin (FITC-BSA) with initial concentration of 0.002 ng/ml is enriched and the concentration factor of a hundred million fold is achieved. And then the technique is implemented to improve the MBAAI sensitivity. In this experiment, magnetic beads coupled with sheep anti-rabbit IgG (immunoglobulin G) are firstly trapped into the immunoassay region via magnetic force, then the FITC-labeled rabbit IgG is electrokinetically concentrated and spontaneously bonded with the previous fixed magnetic beads. After the incubation, the total fluorescence intensity in immunoassay region is calculated. The result shows that the fluorescence intensity is improved by 58 % and the antigen-concentrating with the nanofluidic device can effectively enhance the sensitivity of MBAAI.