Y6DKWcPqtWHBWHlVoWP0fJyQrfFCkGM15O2G9G8u2p241JZak7x0wtH0anSj
Current position: Home >> Scientific Research >> Paper Publications

Study on a method of dynamic response function for the piezoelectric measurement system

Release Time:2019-03-12  Hits:

Indexed by: Journal Article

Date of Publication: 2013-01-01

Journal: Sensors and Transducers

Included Journals: Scopus、EI

Volume: 153

Issue: 6

Page Number: 111-117

ISSN: 17265479

Abstract: Some important areas use more and more piezoelectric force test systems, due to their high sensitivity and natural frequency. To study the dynamic testing performance of the system in the conditions of high natural frequency, one of the effective methods is establishing the mathematical model of dynamic response of test system. By improving specific algorithm of the conventional nonlinear modal fitting method, algorithm of transfer function which is aiming at this test system is deduced, and the transfer function of dynamic response is obtained by using amplitude and phase frequency response curve which is measured by using a piezoelectric test device and the curve is compared with the measured amplitude and phase frequency curve. The rise time, peak time and the overshoot of the step response of the test apparatus is obtained with the inverse Laplace transform. This paper provides a basis of objective evaluation for the dynamic performance of the test device. ? 2013 by IFSA.

Prev One:Piezoelectric sensor of control surface hinge moment

Next One:Research on the dynamic error and acceleration compensation for the piezoelectric sensor