个人信息Personal Information
副教授
硕士生导师
性别:女
毕业院校:大连理工大学
学位:博士
所在单位:化学学院
学科:有机化学
电子邮箱:liufengyu@dlut.edu.cn
Determination of Hg2+ in Tap Water Based on the Electrochemiluminescence of Ru(phen)(3)(2+) and Thymine at Bare and Graphene Oxide-Modified Glassy Carbon Electrodes
点击次数:
论文类型:期刊论文
发表时间:2016-12-01
发表刊物:CHINESE JOURNAL OF CHEMISTRY
收录刊物:SCIE、CSCD、Scopus
卷号:34
期号:12
页面范围:1297-1303
ISSN号:1001-604X
关键字:electrochemiluminescence (ECL); determination; Hg2+; thymine; graphene oxide
摘要:The electrochemiluminescence (ECL) of the Ru(phen)(3)(2+) /thymine (T) system at bare and graphene oxide (GO)-modified glassy carbon (GC) electrodes was utilized to determine Hg2+ in tap water. The ECL intensity of Ru(phen)(3)(2+) was considerably enhanced by the addition of thymine because of the occurrence of ECL tion between them. Subsequently, the ECL intensity of Ru(phen)(3)(2+) /T system rapidly decreased with the addition of Hg2+ because of the formation of a T-Hg2+-T complex. A linear response (R-2=0.9914) was obtained over a Hg2+ concentration range of 1.0x10(-9) mol/L to 1.0x10(-5) mol/L with a detection limit of 3.4x10(-10) mol/L at a bare GC electrode in 0.1 mol/L phosphate buffer (pH=8.0). The detection limit can be further reduced to 4.2x10(-12) mol/L after modification of the GC electrode by GO. To verify its applicability, the proposed method was utilized to determine Hg2+ in tap water and simulated wastewater. The method exhibited good reproducibility and stability and thus reveals the possibility of developing a novel ECL detection method for Hg2+.