Self-Reference Surface Plasmon Resonance Biosensor Based on Multiple-Beam Interference
点击次数:
论文类型: 期刊论文
发表时间: 2016-11-01
发表刊物: IEEE SENSORS JOURNAL
收录刊物: SCIE、EI、Scopus
卷号: 16
期号: 21
页面范围: 7568-7571
ISSN号: 1530-437X
关键字: Surface plasmon resonance; fiber-optic; biosensor; self-reference; multiple-beam interference
摘要: We demonstrated a novel self-reference surface plasmon resonance (SPR) fiber biosensor, which provided a multiple-beam interference referencing signal for refractive index (RI) compensating. The sensor was fabricated by splicing a capillary to a multimode fiber (MMF) coated with a gold film. The gilded MMF acts as the measuring channel while the capillary is used as referencing channel. The experiment result showed that the measuring signal has an irreversible change corresponding specific binding and referencing signal has reversible change. It indicated that the measuring signal can inspect biomolecular interactions in real time and the referencing signal can be used to compensate for interference effects due to bulk RI changes. The sensor provides a high sensitivity of 1470.291 nm/RIU and a resolution about 3.536 x 10(-5) RIU. In addition, this self-reference SPR biosensor we proposed is low cost, simple, and reproduction, which can be applied in biochemical sensing field.