个人信息Personal Information
副教授
硕士生导师
性别:男
毕业院校:大连理工大学
学位:博士
所在单位:光电工程与仪器科学学院
学科:光学工程
办公地点:物理学院233房间
联系方式:xlzhou@dlut.edu.cn
电子邮箱:xlzhou@dlut.edu.cn
Simultaneous measurement of down-hole pressure and distributed temperature with a single fiber
点击次数:
论文类型:期刊论文
发表时间:2012-08-01
发表刊物:MEASUREMENT SCIENCE AND TECHNOLOGY
收录刊物:SCIE、EI、Scopus
卷号:23
期号:8
ISSN号:0957-0233
关键字:Fabry-Perot interferometer; distributed temperature sensor; wavelength-division multiplexing; down-hole monitoring
摘要:A cost-effective sensing system combining the fiber Fabry-Perot interferometer-based pressure sensor with the distributed temperature sensor is presented. This sensing system can realize accurate pressure measurement of the oil reservoir and temperature profile measurement of the wellbore simultaneously. For the sensing wave bands of the pressure and temperature sensor are different, the wavelength-division multiplexing (WDM) technology was used to couple the two sensing signals into one single fiber. As the wavelength difference between Stokes peak and anti-Stokes peak is larger than the bandwidth of commercial WDM, a wide bandwidth WDM with gain flattening range up to 80 nm was custom-made. Experimental results showed that this sensing system can reach 0.1% accuracy for the pressure measurement with a 0-30 MPa working range and 1 degrees C accuracy for the temperature measurement in a working range from room temperature to 300 degrees C. Additionally, a resolution of below 1.4 kPa for the pressure measurement and 0.5 degrees C for the temperature measurement had been realized. Field test results showed that this sensing system also has good long-term stability in high-temperature environment down hole.