location: Current position: Home >> Scientific Research >> Paper Publications

Automatic emissive probe apparatus for accurate plasma and vacuum space potential measurements

Hits:

Indexed by:期刊论文

Date of Publication:2018-02-01

Journal:PLASMA SCIENCE & TECHNOLOGY

Included Journals:SCIE、EI、CPCI-S

Volume:20

Issue:2

ISSN No.:1009-0630

Key Words:space potential; emissive probe; probe heating current; LabVIEW

Abstract:We have developed an automatic emissive probe apparatus based on the improved inflection point method of the emissive probe for accurate measurements of both plasma potential and vacuum space potential. The apparatus consists of a computer controlled data acquisition card, a working circuit composed by a biasing unit and a heating unit, as well as an emissive probe. With the set parameters of the probe scanning bias, the probe heating current and the fitting range, the apparatus can automatically execute the improved inflection point method and give the measured result. The validity of the automatic emissive probe apparatus is demonstrated in a test measurement of vacuum potential distribution between two parallel plates, showing an excellent accuracy of 0.1 V. Plasma potential was also measured, exhibiting high efficiency and convenient use of the apparatus for space potential measurements.

Pre One:Phase shift effects of radio-frequency bias on ion energy distribution in continuous wave and pulse modulated inductively coupled plasmas

Next One:Experimental investigation of the electron impact excitation behavior in pulse-modulated radio frequency Ar/O2 inductively coupled plasma