location: Current position: Home >> Scientific Research >> Paper Publications

Topology optimization based design of unilateral NMR for generating a remote homogeneous field

Hits:

Indexed by:期刊论文

Date of Publication:2017-06-01

Journal:JOURNAL OF MAGNETIC RESONANCE

Included Journals:SCIE、EI、PubMed

Volume:279

Page Number:51-59

ISSN No.:1090-7807

Key Words:Topology optimization; Structural design; Unilateral NMR; Homogeneous field

Abstract:This paper presents a topology optimization based design method for the design of unilateral nuclear magnetic resonance (NMR), with which a remote homogeneous field can be obtained. The topology optimization is actualized by seeking out the optimal layout of ferromagnetic materials within a given design domain. The design objective is defined as generating a sensitive magnetic field with optimal homogeneity and maximal field strength within a required region of interest (ROI). The sensitivity of the objective function with respect to the design variables is derived and the method for solving the optimization problem is presented. A design example is provided to illustrate the utility of the design method, specifically the ability to improve the quality of the magnetic field over the required ROI by determining the optimal structural topology for the ferromagnetic poles. Both in simulations and experiments, the sensitive region of the magnetic field achieves about 2 times larger than that of the reference design, validating validates the feasibility of the design method. (C) 2017 Published by Elsevier Inc.

Pre One:Method to Further Improve Sensitivity for High-Order Vibration Mode Mass Sensors With Stepped Cantilevers