location: Current position: Home >> Scientific Research >> Paper Publications

Phase difference signal processing technology in surface plasmon resonance sensing system

Hits:

Indexed by:期刊论文

Date of Publication:2015-06-10

Journal:Zhongguo Jiguang/Chinese Journal of Lasers

Included Journals:EI、PKU、ISTIC、Scopus

Volume:42

Issue:6

ISSN No.:02587025

Abstract:In order to improve the resolution of phase modulated surface plasmon resonance (SPR) measurement technology. The signal processing technology is optimized. In the differential phase SPR biosensor based on Mach-Zehnder configuration, the SPR phase can be extracted by comparing the phase difference between the interference signals of p-polarization and the reference signals of s-polarization. The 2-path interference optical signal is recevied by computer. According to the mathematical theory and signal processing knowledge, select and processing the interference optical signals. The phase difference of 2-path signal is analyzed at the same time. In the signal processing technology, the phase difference of ups and downs is 0.006  , and the sensitivity limit is 4.7  10-7 RIU. ?, 2015, Science Press. All right reserved.

Pre One:基于光导毛细管器件的 SPR生物传感器及 IgG检测

Next One:Narrow-band wavelength tunable filter based on asymmetric double layer metallic grating