应用于激光诱导荧光检测的微透镜阵列
点击次数:
论文类型:期刊论文
发表时间:2022-06-30
发表刊物:激光与光电子学进展
卷号:54
期号:8
页面范围:080402
ISSN号:1006-4125
摘要:Microfluidic systems based on laser induced fluorescence(LIF)detection
technology have been widely used in the field of biochemistry
detection.Aiming at the problem of weak fluorescence intensity induced
by a few detection samples in the microfluidic system,a microfluidic
chip integrating a microlens array(MLA)is designed and fabricated to
improve the intensity of the detected fluorescence.An 8*8photoresist MLA
mold with the diameter variation coefficient of 0.36%is obtained by the
hot melt technology.Then,a cover plate integrating a
polydimethylsiloxane MLA is produced by the soft-lithography
technology,and the focal length uniformity error is 7%.Afterwards,a
micro-channel substrate is made.The substrate and the cover plate are
packaged by the oxygen plasma bonding technology.Finally,the
fluorescence intensity of the chip is detected by a fluorescence
microscope, and the chip is injected by fluorescein isothiocyanate
fluorescent dye solution with concentration of 10mumol·L~(-1). The
result shows that the fluorescence intensity with the microlens is
improved about 2.2times as much as the fluorescence intensity without
the microlens.
备注:新增回溯数据